Section 333-122-0110 - General Requirements for All RGD: Surveys
Universal Citation: OR Admin Rules 333-122-0110
Current through Register Vol. 63, No. 12, December 1, 2024
(1) All X-ray systems shall have radiation surveys performed, as required by OAR 333-120-0200:
(a) Upon installation of the equipment;
(b) Following any change in the
initial arrangement, number or type of local components in the system;
(c) Following any maintenance
requiring the disassembly or removal of a local component in the system;
(d) During the performance of
maintenance and alignment procedures if the procedures require the presence of
a primary X-ray beam when any local component in the system is disassembled or
removed;
(e) Any time a visual
inspection of the local components in the system reveals an abnormal condition;
and
(f) Whenever personnel
monitoring devices show a significant increase over the previous monitoring
period or the readings are approaching the limits specified in OAR
333-120-0100.
(2) Radiation levels must not exceed the limits outlined in OAR 333-120-0180.
(3) The registrant must conduct all surveys with a calibrated and operable radiation survey instrument.
Statutory/Other Authority: ORS 453.605 - 453.807
Statutes/Other Implemented: ORS 453.605 - 453.807
Disclaimer: These regulations may not be the most recent version. Oregon may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.